Ultra modern design with a user friendly interface
The water chamber with drawer structure and opening cover, is more convenient to fill water and clean.
3-button control panel with clear user interface makes it simple and easy to use.
Intelligent comfort features
Inspiratory / Expiratory sensitivity
8 grades of I Sens and E Sens optimize the compliance of the device with the patient.
Newly designed integrated humidifier has large heating plate (> 50 cm2), which ensures excellent humidifying capacity and improves comfort.
The ramp time ranges from 0 to 60 minutes
Efficiencies in Therapy
Auto CPAP pressure automatically to improve patient comfort based on monitoring of apnea, hypopnea and snoring events.
AutoS mode
Auto adjusting S mode, according to the sleep events, the pressure changes automatically within the set range.
Automatic leakage compensation ensures the therapy accuracy anywhere.
S/T Mode
The device complies with patient‘s breathing. However, if there is no inhalation for a certain period of time, the device will give a forced ventilation to ensure the minimum ventilation.
Target Tidal Volume Function
√ Optimize IPAP according to mean Vt of last 5 breathings and prescribed Target Vt. Larger difference between mean Vt and target Vt takes more evident adjustment in IPAP.
√ Inspiratory pressure is between IPAP min and IPAP max.
√ Larger difference between mean Vt and target Vt takes more evident adjustment in IPAP.
Patient Vt is the mean of Vt values from patient's last 5 breathings.
Powerful data transfer/retrieval solution
Various way of therapy report review
Quick Report through Device Screen
PAP Link App
PAP Link Web
PAP Link PC
Complete optional accessories
G2S adopts a series of complete accessories to meet different usage scenarios and needs:
SpO2 Kit (adapted for G2S BPAP System series)
Cellular Module
WiFi Module
Heated Tubing

Specifications
Device Size
Dimensions: 274 mm × 184 mm × 115 mm
Weight: 1.9 kg
Heated Humidifier
Humidifier Settings: Off, 1 to 5 (95°F to 154.4°F / 35°C to 68°C)
Humidifier Output: No less than 10 mg H2O/L
Water Capacity
To maximum fill line 360 mL
SD Card
The SD card can record patient data and fault information
AC Power Consumption
100 - 240 V AC, 50/60 Hz, 2.0 A
Sound Pressure Level
< 30 dB, when the device is working at the pressure of 10 hPa
Automatic leak compensation
The pressure error ≤ ± 1 hPa when the air leakage is no more than 70 L/min.
Maximum Delivered Gas Temperature
≤ 43°C
Static Pressure Stability
± 0.5 hPa
SpO2
0% to 100%
Pulse Rate
40 to 240 BPM